[vc_section][vc_row][vc_column][vc_custom_heading text=”QUICK IMPLANT SIZE IDENTIFICATION” font_container=”tag:h2|text_align:center” use_theme_fonts=”yes”][vc_empty_space][vc_column_text]
[/vc_column_text][vc_row_inner][vc_column_inner width=”1/6″][vc_column_text]
| REGULAR PALTFORM ( REDICE ACTIVE ) | |||||
| Length/Diameter | Ø 3.3 | Ø 3.75 | Ø 4.20 | Ø 5.0 | Ø 6.0 |
| 6.0mm L | RA37506 | RA4206 | RA5006 | RA6006 | |
| 8.0mm L | RA3308 | RA37508 | RA4208 | RA5008 | RA6008 |
| 10.0mm L | RA3310 | RA37510 | RA4210 | RA5010 | RA6010 |
| 11.5mm L | RA33115 | RA375115 | RA42115 | RA50115 | RA60115 |
| 13.0mm L | RA3313 | RA37513 | RA4213 | RA5013 | RA6013 |
| 16.0mm L | RA3316 | RA37516 | RA4216 | RA5016 | RA6016 |
Hex 2.42
[/vc_column_text][vc_single_image image=”10858″ alignment=”center”][/vc_column_inner][vc_column_inner width=”1/6″][vc_separator color=”orange” border_width=”5″][vc_single_image image=”10859″][/vc_column_inner][vc_column_inner width=”1/6″][vc_separator color=”juicy_pink” border_width=”5″][vc_single_image image=”10860″][/vc_column_inner][vc_column_inner width=”1/6″][vc_separator color=”blue” border_width=”5″][vc_single_image image=”10861″][/vc_column_inner][vc_column_inner width=”1/6″][vc_separator color=”green” border_width=”5″][vc_single_image image=”10862″][/vc_column_inner][vc_column_inner width=”1/6″][vc_separator color=”mulled_wine” border_width=”5″][vc_single_image image=”10863″][/vc_column_inner][/vc_row_inner][/vc_column][/vc_row][vc_row][vc_column][vc_empty_space height=”60px”][vc_row_inner][vc_column_inner width=”1/2″][vc_column_text]SINGLE PROSTHETICAL PLATFORM
The RDI Internal hex diameters utilize one line of prosthetic components.[/vc_column_text][/vc_column_inner][vc_column_inner width=”1/2″][vc_single_image image=”11350″ img_size=”full” alignment=”center”][/vc_column_inner][/vc_row_inner][/vc_column][/vc_row][vc_row][vc_column][vc_separator][vc_custom_heading text=”DRILLING PROTOCOL” use_theme_fonts=”yes”][vc_empty_space][vc_row_inner][vc_column_inner width=”1/4″][vc_column_text]
3.3
[/vc_column_text][/vc_column_inner][vc_column_inner width=”1/4″][vc_single_image image=”11089″ img_size=”full” alignment=”center”][/vc_column_inner][vc_column_inner width=”1/2″][vc_column_text]- Drill sequence by bone type – An additional 1 mm must
be added to the length of the drill to account for the angled
cutting tip. - D-I, D-II Dense bone considerations – Countersink
should be used if the standard dense bone protocol is not
sufficient to fully seat the implant without exceeding the
recommended maximum insertion torque of max. 60Ncm
for implant Ø3.75mm and above. Do not exceed max.
45Ncm for implants diameter lower than 3.75mm.
3.75
[/vc_column_text][/vc_column_inner][vc_column_inner width=”1/4″][vc_single_image image=”11090″ img_size=”full” alignment=”center”][/vc_column_inner][vc_column_inner width=”1/2″][vc_column_text]- Drill sequence by bone type – An additional 1 mm must
be added to the length of the drill to account for the angled
cutting tip. - D-I, D-II Dense bone considerations – Countersink
should be used if the standard dense bone protocol is not
sufficient to fully seat the implant without exceeding the
recommended maximum insertion torque of max. 60Ncm
for implant Ø3.75mm and above. Do not exceed max.
45Ncm for implants diameter lower than 3.75mm.
4.2
[/vc_column_text][/vc_column_inner][vc_column_inner width=”1/4″][vc_single_image image=”11091″ img_size=”full” alignment=”center”][/vc_column_inner][vc_column_inner width=”1/2″][vc_column_text]- Drill sequence by bone type – An additional 1 mm must
be added to the length of the drill to account for the angled
cutting tip. - D-I, D-II Dense bone considerations – Countersink
should be used if the standard dense bone protocol is not
sufficient to fully seat the implant without exceeding the
recommended maximum insertion torque of max. 60Ncm
for implant Ø3.75mm and above. Do not exceed max.
45Ncm for implants diameter lower than 3.75mm.
5.0
[/vc_column_text][/vc_column_inner][vc_column_inner width=”1/4″][vc_single_image image=”11092″ img_size=”full” alignment=”center”][/vc_column_inner][vc_column_inner width=”1/2″][vc_column_text]- Drill sequence by bone type – An additional 1 mm must
be added to the length of the drill to account for the angled
cutting tip. - D-I, D-II Dense bone considerations – Countersink
should be used if the standard dense bone protocol is not
sufficient to fully seat the implant without exceeding the
recommended maximum insertion torque of max. 60Ncm
for implant Ø3.75mm and above. Do not exceed max.
45Ncm for implants diameter lower than 3.75mm.
6.0
[/vc_column_text][/vc_column_inner][vc_column_inner width=”1/4″][vc_single_image image=”11094″ img_size=”full” alignment=”center”][/vc_column_inner][vc_column_inner width=”1/2″][vc_column_text]- Drill sequence by bone type – An additional 1 mm must
be added to the length of the drill to account for the angled
cutting tip. - D-I, D-II Dense bone considerations – Countersink
should be used if the standard dense bone protocol is not
sufficient to fully seat the implant without exceeding the
recommended maximum insertion torque of max. 60Ncm
for implant Ø3.75mm and above. Do not exceed max.
45Ncm for implants diameter lower than 3.75mm.